PCE-HLD 10 - Hydrogen Detector

by Pce
 Pce
Κωδικός Τιμή ** Στοκ
PCE-HLD 10 €420,00 2-5 εργάσιμες


Τιμές χωρίς ΦΠΑ 24%


High sensitivity >2 g (0.07 oz) / year
Average sensitivity >15 g (0.53 oz) / year
Low sensitivity >30 g (1.05 oz) / year
Sensitivity refers to 5% Hydrogen (H2) and 95% Nitrogen (N2)

Alarm Audio and visual
Heating 45 seconds
Battery life ~12 hours
Power supply 4 x 1.5V AA batteries
Automatic shutdown 10 minutes
Maximum altitude 2000 m / 6561 ft asl
Operating conditions 0 ... 40 °C / 32 ... 104 °F, <80 % RH, non-condensing
Storage conditions -10 ... 60 °C / 14 ... 140 °F, <70 % RH, non-condensing
Pipe length 40cm / 15.7"
Dimensions 213 x 65.5 x 53.5mm / 83.8 x 25.7 x 21.0"
Weight 400g / 14.1oz

Gas detector
With audible and visual alarm / 40 cm (15.7") long semi-rigid tube /
Adjustable sensitivity levels / Detection starts at less than 5 ppm


With the selective hydrogen sensor, the gas detector can detect even the smallest leaks. Upon activation, the device automatically records the current hydrogen concentration in the environment and then only evaluates changes compared to the original measurement. 
In case the measurement is not clear with all LED lights on, the current hydrogen concentration can be set as the base value by pressing the reset button and then the leak can be detected more accurately. After taking a measurement, the detector can be reset to previous sensitivity levels by pressing the RESET button.

The user first selects the sensitivity level by pressing the corresponding button.
 "H" corresponds to high sensitivity, "M" to medium and "L" to low. The hydrogen content of air under normal conditions is 0.5 ppm. The meter detects the minimum concentration of 5 ppm, i.e. 0.0005 %.

The sensor is mounted on the head of a 40 cm long semi-rigid tube, so that it can easily reach hard-to-reach places. If the gas detector detects a leak, an alarm sounds and additional LED indicators light up. The higher the recorded hydrogen concentration compared to the base level, the more indicator lights light up.
The device is powered by four batteries that provide an autonomy of twelve hours of normal use. After 10 minutes of inactivity, it turns off to save power. The built-in detection sensor has a limited lifetime. Under normal conditions of use, it can be used for a period of more than one year. The correct operation of the sensor can be checked by regular use of test gas. If the sensor checks and turns out to be unreliable, it can be replaced with a new one.

- Bar graph with color gradient (green, orange, red)
- Three adjustable sensitivity levels
- Battery life approx 12 hours
- Semi-rigid sensor with a length of 40 cm / 15.7"
- Audible and visual alarm
- Auto power off
Package contents
1 x PCE-HLD 10 Gas Detector
1 x Carrying case
4 x 1.5V AA batteries
1 x User Manual