Vickers Hardness Measurement Method

The Vickers method is a static hardness testing method, which has the following characteristics:·

  • It is a hardness measurement method specified by ISO 6507, ASTM E92, ASTM E384 standards.
  • The Vickers method has a test load range from 1 gf to 120 kgf according to ASTM, which means that this method can be used for hardness tests across all load scales
  • It is an optical method. This means that the size of the indentation left by the indenter on the specimen is measured to determine the hardness value.
  • Indenter shape and material: The indenter is a diamond equilateral pyramid (with a square base) with an included angle of 136°.

TABLE 1

Load range Method Indenter Test force F Applications  
Micro hardness range HV0.01 Diamond pyramid 136ο 0.098Ν Structural components, very thin layers Range 1HV- 3000HV
HV0.025 0.245Ν
HV0.05 0.490Ν
HV0.1 0.980Ν
Low-force range HV0.2 Diamond pyramid 136ο 1.961Ν Thin layers CHD/SHD/NHD, thin metal sheets, small specimens
HV0.3 2.942Ν
HV0.5 4.903Ν
HV1 9.807Ν
HV2 19.61Ν
HV3 29.42Ν
Macro range (common hardness range) HV5 Diamond pyramid 136ο 49.03Ν Standard specimens
HV10 98.07Ν
HV20 196.1Ν
HV30 294.2Ν
HV50 490.3Ν
HV100 980.7Ν

Operating principle

Vickers hardness (HV) is obtained from the quotient of the applied test force (F in newton N) and the surface area of the indentation on the sample (Figure 1). To calculate the indentation surface area, the average of the two diagonals (d1 and d2 in mm) is used, because the base area of Vickers indentations is often not exactly square. Ideally, the test force is increased from 0 to the final value within 7 seconds (minimizing measurement uncertainty). The maximum permissible interval for the application time according to the standard is 2 to 8 seconds (nominal duration 7 seconds). In general, the dwell time for the test force is 10 to 15 seconds (nominal duration 14 seconds).

Operating principle of Vicker hardness measurement.

Figure 1: Operating principle of Vicker hardness measurement.

Sample requirements for the Vickers method

When the Vickers hardness testing method is used, it is usually necessary to prepare the surface of the sample to be tested, as the sample surface quality requirements are significantly stricter than those of the Rockwell method, where it is sufficient for the surface to be free of dust and impurities.

:::info
Abbreviation explanation

1. The numerical hardness value (between 1 and 3000).

2. The two letters "HV" indicate the Hardness Vickers method.

3. The applied test load in kgf. (Table 1).