Operating frequency | 60 kHz, sine wave |
Conductivity measurement range |
0.51% IACS ... 112% IACS 0.3MS/m ... 65MS/m |
Resistance range | 0.015388 ... 3.33333 Ω • mm² / m |
Conductivity measurement analysis |
0.01% IACS (at <51% IACS) 0.1% IACS (at 51% IACS ... 112% IACS) |
Accuracy of conductivity measurement |
± 0.5% at 20°C / 68°F ± 1% in the range 0 ... 40°C / 32 ... 104°F |
Withdrawal of compensation | 0.5 mm / ≈ 0.0197 in |
Temperature measurement range | 0 ... 50°C / 32 ... 122°F |
Temperature measurement accuracy | ±0.5° |
Temperature compensation | Automatic adjustment of the conductivity result to the value at 20°C / 68°F |
Operating conditions | 0 ... 50°C / 32 ... 122°F, 0 ... 95% RH |
Screen | LCD, with backlight |
Menu languages | English, German, Chinese (Simplified) |
Power supply | Internal, rechargeable battery |
Specimen diameter | 14 mm / ≈ 0.55 in |
Memory | Stores up to 500 groups of measurement values |
Exit | USB |
Dimension | ≈ 220 x 95 x 35 mm / 8.66 x 3.74 x 1.38 in |
Bar | ≈ 415 g / 1 lb (with sample) |
PCE-COM 20 Conductivity meter for non-ferrous metals
Large measurement range up to 112% IACS or 65 MS/m with elevation and temperature compensation, built-in memory, PC compatible software, 60 kHz operating frequency
The PCE-COM 20 is a portable non-destructive conductivity meter for non-ferrous metals. This non-destructive testing (NDT) device uses eddy current to determine the electrical conductivity of non-ferrous metals such as aluminum and copper.
Featuring an operating frequency of 60 kHz, the conductivity meter has a wide measuring range of 0.51 ... 112% IACS with an accuracy of +/- 0.5% at 20°C / 68°F. Thanks to a reading of up to 0.01% IACS, automatic temperature compensation (ATC) and lift compensation (up to 500 µm), the device ensures fast, easy, accurate and reproducible measurements of coated or corroded parts during changing environmental conditions.
Measuring the electrical conductivity of non-ferrous metals, such as aluminium, copper, titanium, magnesium or bronze, is essential in applications such as evaluating the tolerance limits of heat-treated or hardened materials, checking the authenticity of coin alloys, verifying alloy inhomogeneity in critical components and the classification of materials.
The conductivity meter is powered by an internal battery that can be charged via the interface on the bottom of the unit. In addition to the connection for the charging adapter, there is a USB port that allows the transfer of measurement data to a computer for more detailed analysis and processing of the data.
- For non-ferrous metals and alloys
- Non Destructive Testing (NDT)
- Operating frequency 60 kHz
- Large measuring range up to 112% IACS or 65 MS/m
- Memory for up to 500 measurement groups
- Automatic Temperature Compensation (ATC)
- Elevation compensation up to 500 µm
- Automatic calibration
- Adjustable backlight
- Internal rechargeable battery
- Portable use
- Titanium 1.03% IACS, Copper 8.11% IACS and Copper 100% IACS samples
1 x Sine-wave measuring sensor
1 x Set of Calibration Plates (Titanium 1.03% IACS, Bronze 8.11% IACS, and Copper 100% IACS)
1 x PC compatible software
1 x USB to RS-232 data cable
1 x Charging Adapter
1 x Hex Wrench
1 x Carrying case
1 x User Manual