CODEPCE-MND 10
CODEPCE-MND 10
Noise Dosimeter - Data Logger Class II
- IEC 61252:2002, Class 2
- Range 70 ... 140 dB
- Battery life >15 hours
- Weighting A, B and Z
- USB readout with software
- Optional ISO calibration certificate
Availability & price
| Code | Availability | excl. VAT | incl. VAT | Certificate | Cart |
|---|---|---|---|---|---|
| PCE-MND 10 |
€1.233,00
|
€1.528,92
|
- IEC 61252:2002, Class 2
- Range 70 ... 140 dB
- Battery life >15 hours
- Weighting A, B and Z
- USB readout with software
- Optional ISO calibration certificate
The PCE-MND 10 is a badge-type noise dose meter for personal noise exposure at work, to IEC 61252:2002. It measures 70 ... 140 dB with frequency weighting A, B and Z, and time weighting fast, slow or impulse. The 470 mAh battery lasts over 15 hours.
Up to three virtual dose meters run in parallel, so three measurement standards are recorded at once. The display shows up to 14 individually configurable measurement windows. Data goes to the PC via the charging station; the software reads out, analyses and configures the device. Coupled charging stations charge up to four dosimeters from one power supply. Two crocodile clips attach it to the clothing, 10 to 15 cm (3.9 and 5.9") from the ear.
Highlights
- Badge type, IEC 61252:2002 and ANSI S1.25-1991
- Measuring range 70 ... 140 dB
- Frequency weighting A, B and Z
- Time weighting fast, slow or impulse
- Battery life over 15 hours
- Three virtual dose meters at once
- Mini-USB interface and key lock
- Optional ISO calibration certificate
| Sound level | |
|---|---|
| Measuring range | 70 ... 140 dB |
| Resolution | 0,1 dB |
| Accuracy | <2.2 dB at 94 dB and 1 kHz |
| Measuring range | 90 ... 140 dBZ |
| Resolution | 0,1 dB |
| Accuracy | <2.2 dB at 94 dB and 1 kHz |
| General technical data | |
| Measuring functions | LXPeak, LXE, LXeq, LXY(SPL), ALARM, MAX, MIN, Sepa²h, PLEX8H, LEX8H, LAS, PLEPd, LEPd, PTWA, TWA, PDOSE, Dose, LAVG |
| Measuring functions (additional information) | X = Frequency weighting: A, B, C, Z; Y = time weighting: F, S, I; Many other measurement functions (see user manual) |
| Time weighting | Pulse (35ms), slow (1 s), fast (125 ms) |
| Frequency weighting | Z, C, A |
| Frequency | 31.5 ... 8000 Hz |
| Display type | LCD |
| Storage capacity | 6101 Data records |
| Storage interval from | 60 s |
| Storage interval to | 120 min. |
| Microphone type | Electret capacitor |
| Interface | Mini-USB |
| Standard(s) | Class 2, ANSI S1.25-1991, IEC 61252: 2002 |
| Warm-up time | 10 s |
| Operating time | 15 h |
| Menu language | English, English (GB) |
| Protection class (device) | IP20 |
| Power supply | 100 ... 240V AC | 50/60 Hz | 0.7 A |
| Connector type | Device-europlug |
| (Rechargeable) battery | x 3,7 V internal , Lithium polymer battery |
| Capacity | 470 mAh |
| Operating conditions | 0 ... 50 °C , 10 ... 90 % RH |
| Storage conditions | -10 ... 60 °C , 10 ... 75 % RH |
| Dimensions ( L x W x H ) | 85 x 50 x 55 mm |
| Weight | 105 g |
- 1×Sound Dosimeter (badge type) PCE-MND 10
- 1×USB cable
- 1×Charging station with power adapter
- 1×Supply bridge
- 1×Mounting bridge
- 1×Carrying case
- 1×Instruction manual
- Software download
- LINKS:
- ========================================
- /english/english/download-win_4.htm
Official distributor
PCE Instruments
Sound Level Meters
You're viewing one product from this category. Browse the full range.