CO2 - O2 Gas Detector
PCE-WMM 100
| Measurement function | CO2 |
| Measurement range | 0 .... 50000 ppm |
|
Reading |
<10000 ppm: 10 ppm |
| >10001 ppm: 100 ppm | |
| Accuracy | ± 100 ppm / 5 % of the measured value |
| Measurement function | O2 |
| Measurement range | O2: 0 ... 30 % |
|
Reading |
0.1% |
| Accuracy | < 3 % of the measured value |
| Measurement function | Temperature |
| Measurement range | 0 ... 50 °C / 32 ... 122 °F |
|
Reading |
0.1 °C |
| Accuracy | ± 1 °C |
| Repeatability | CO2: ±20 ppm at 400 ppm |
| O2: < ±1 % | |
| Temperature dependence | ± 0.3 % of measured value / °C |
| (at 25 °C) | ± 2 ppm / °C |
| (higher price applies) | |
| Dependence on pressure | 0.13 % of the measured value / mmHg |
| Response time | CO2: T90 < 60s |
| O2: T90 < 30s | |
| Temperature: 20 ... 30 m | |
| Alarm 1 | CO2: 5000 ppm, 1 %, 1.5 %, 2 % |
| O2: 18 %, 18.5 %, 19 %, 19.5 %, 20 % | |
| Alarm 2 | CO2: 1.5%, 2%, 2.5%, 3%, 3.5%, 4% |
| O2: 16 %, 16.5 %, 17 %, 17.5 %, 18 % | |
| Alarm volume | 80 db / 0.1 m |
| Warm-up time | < 60 s at an ambient temperature of 22 °C |
| Power supply | Current transformer 100 ... 240 V AC / 6 V DC / 0.5 A |
| Analog output | 2 x 4 ... 20mA |
| Load | <150 Ohms |
| Relay | Maximum 2 A / 30 V DC / 250 V AC |
|
Operating conditions |
0 ... 40 °C / 32 ... 104 °F / max. 95 % rH |
| Storage conditions | -20 ... 60 °C / -4 ... 140 °F / max. 95% rH |
| Sealing class | Main unit: IP54 |
| Remote Unit:- | |
| Dimensions | Main unit: 170 x 63 x 26 mm / 6.6 x 2.4 x 1.0" |
| Remote unit: 80 x 35.7 x 10 mm / 3.1 x 1.4 x 0.3" | |
| Weight | Main unit: 1200g with cable |
| Remote unit: 120g / 4.2oz |
1 x Remote unit
1 x 8m connection cable
1 x Power Adapter
1 x Mounting kit
1 x User Manual