CODEPCE-AS1
CODEPCE-AS1
Air Quality Meter for Microbiological Contamination ISO 14698-1 (Hospitals - Laboratories etc.)
- 100 lpm air sampling
- To ISO 14698-1
- Microbiological air monitoring
- Portable, battery-powered device
Availability & price
| Code | Availability | excl. VAT | incl. VAT | Certificate | Cart |
|---|---|---|---|---|---|
| PCE-AS1 |
€3.915,00
|
€4.854,60
|
- 100 lpm air sampling
- To ISO 14698-1
- Microbiological air monitoring
- Portable, battery-powered device
The PCE-AS1 is a portable air sampler for assessing microbiological contamination and monitoring hygiene. It samples airborne microbes, bacteria, mould, fungi and viruses at 100 lpm. It is built for hospitals, clean rooms, and food and beverage, cosmetics and pharmaceutical production. It meets current requirements for airborne microbial analysis. An ISO calibration lab certificate is available as an option.
It works on the aerosol impaction (Andersen) principle. Air drawn in at the top passes through 300 micropores. The laminar flow hits the surface of an agar petri dish. After sampling, you remove and incubate the dish, then assess bacterial and fungal counts to USP 797 and CAG-009 guidance. Sterilise the collecting head in an autoclave at 121 °C for 30 minutes, or wipe it with a cloth soaked in 75% alcohol. Sterilise the petri dish with 75% alcohol.
Highlights
- Backlit LCD
- Memory for up to 99 parameters
- Collecting head sterilisable by autoclave or alcohol
- Adjustable sample volume up to 6000 l
- Battery with 6400 mAh capacity
- Takes standard petri dishes
| General technical data | |
|---|---|
| Display type | LCD with backlight |
| Display size | 3,5 Inch |
| Storage medium | Internal memory |
| Storage capacity | 2000 Data records |
| Memory capacity (additional information) | Memory for up to 99 individual parameters. |
| Standard(s) | ISO 14698-1 |
| Flow rate | 0,38 m/s |
| Alarm | Acoustic |
| Sample quantity | 10 ... 6000 l |
| Micropores in the collection head | 300 |
| Sample throughput | 100 l/min |
| Bore diameter | 0,6 mm |
| Insertable culture tray | Ø70 ... 90 x 15 mm |
| Rehearsal runs | 99 runs with up to 100 hours running time |
| Menu language | English, English (GB), Chinese |
| Protection class (device) | IP20 |
| Power supply | 100 ... 240V | AC 50/60 Hz |
| Connector type | Device-europlug |
| Capacity | 6400 mAh |
| Operating conditions | 0 ... 35 °C , 10 ... 90 % RH |
| Storage conditions | 0 ... 35 °C , 10 ... 90 % RH |
| Dimensions ( L x W x H ) | ø 120 x 300 mm |
| Weight | 2,6 kg |
- 1×Air Sampler PCE-AS1
- 1×Adapter for glass petri dishes with calibration attachment
- 1×Mains power adapter
- 1×Carrying case
- 1×Instruction manual
- Software free as QR image for IOS / Android
Official distributor
PCE Instruments
Gas Detectors - Dust Meters
You're viewing one product from this category. Browse the full range.